FLUKE 5725A PDF

Other Info: The Fluke A Amplifier is a companion to the A Calibrator. It extends the A’s workload coverage. It increases maximum direct and. Buy the Used Fluke A Calibrator Amplifier for A on sale at Transcat. Free shipping over $ Rent the Fluke A Calibrator Amplifier for A at Weekly and Monthly rates available, request a quote today!.

Author: Vudoran Nigami
Country: Libya
Language: English (Spanish)
Genre: Technology
Published (Last): 6 September 2004
Pages: 389
PDF File Size: 10.79 Mb
ePub File Size: 18.58 Mb
ISBN: 619-1-97889-746-7
Downloads: 54995
Price: Free* [*Free Regsitration Required]
Uploader: Dacage

Fluke 5725A Amplifier

Test Instructions These instruments can produce or measure lethal voltages at their terminals. Model A operating functions and ranges: Since most meters with a high current range use a separate high current input terminal, this configuration normally eliminates the need to move cables during a procedure.

The A from Fluke is a bench top amplifier for the A and A. The A Amplifier enhances the A Series Multifunction Calibrators in the ac voltage, ac current, and dc current functions.

These instruments can produce or measure lethal voltages at their terminals. This can increase internal operation temperatures affecting reliability.

If a single-point current output is needed, the A Series Calibrator can be configured to source all current outputs through the A binding posts. If the instrument is smoking or has a bad smell coming from it turn it off immediately and contact 5725q for further instruction.

You are being redirected

Extended-range ac and dc current is supplied through them. Model Name Description A Amplifier. If the unit has been calibrated within the last two calendar years and fkuke is a large calibration shift tluke, this can be considered a failure as well. This eliminates the need to move cables during a procedure that requires amplified as well as standard calibrator outputs. We’d like your feedback. Register Log in Contact Us. If self tests pass, run cal zero from the calibration menu.

  GIGABYTE GA-965P-DS3 MANUAL PDF

For any functional or calibration errors not detected by the self test and cal operations, contact us for repair. In the worst cases stray leakage caused by dust in the A can cause uncontrolled high voltage conduction within the instrument resulting in and significant damage. It extends the V range alternating voltage function up to kHz and the direct and alternating current functions up to 11 Amps. Fluke Calibration Silver CarePlans are available for most calibration products.

Age — Very old instruments can suffer from age related failures caused by normal deterioration of components and assemblies over time.

There are often multiple errors. Output Tripped To Standby V supply magnitude too high in operate, most likely an external fault.

Rent Fluke A Calibrator Amplifier | Transcat

Testing should only be performed by qualified technicians. Mechanical damage — The A contains several large transformers and is very heavy. Common Failures Poor Maintenance — Over time or in dusty working environments flike and debris can build up on the filter and inside the unit.

  HYPATIA OF ALEXANDRIA MATHEMATICIAN AND MARTYR PDF

Any failure or operational deficiency information available allows us to provide the best service.

Voltage output from the A is available at the A Series Calibrators’ front or rear binding posts. Power up the A and A.

Frequency limits at higher voltage increase to kHz at V, 30 kHz at V Load limits are to 70 mA for frequencies above 5 kHz, and to 50 mA for frequencies less than 5 kHz Capacitive load limits are increased to pF. It can also create leakage paths that affect calibration stability and operation accuracy. Please include any information available about failures and what function the instrument was in prior to failure.

Dropping or other impact damage from transport and shipping is not uncommon. If zero cal passes, run cal check from the calibration menu. Run A instrument self diag from the utilities menu.

Product Manuals A Getting Started Report as many as you reasonably can. Please contact your local Fluke Calibration sales representative for details or to request a quote. The A operates under complete control of the A Series calibrators through an interface cable supplied with the A.